ESTUDIO IN VITRO DE LOS CAMBIOS EN LAS PROPIEDADES SUPERELASTICAS DE LOS RESORTES DE NITINOL EXPUESTOS A RECICLAJE

AN IN VITRO STUDY ABOUT THE CHANGES IN SUPERELASTIC PROPERTIES OF NITINOL COIL SPRING EXPOSED TO RECYCLING


Dr. Marco Aurelio Pardo

Claudia Bonilla González
Cristina Leon Jatem




CORRESPONDENCIA

Marco Aurelio Pardo, Pontificia Universidad Javeriana, Facultad de Odontología, Departamento del Sistema Craneofacial. Carrera 7 # 40-62, edificio 26. Bogotá, D. C., Colombia. Teléfono: +57-1-320.83.20, extensión 2880.

Claudia Bonilla González. bonillac@javeriana.edu.co

Cristina León Jatem. cristinleon@hotmail.com  o cristinleon@scdeo.com

 

TITULILLO: RESORTES DE NITINOL Y CICLOS DE RECICLAJE

RESUMEN

El objetivo fue describir los cambios en las propiedades superelásticas de los resortes de nitinol expuestos a varios ciclos de reciclaje. Se tomaron 12 resortes de Nitinol 3MUnitek® de 9mm de longitud, y se determinó su comportamiento en la curva tensión/deformación (T0). Posteriormente se dividieron en dos grupos, 6 resortes se esterilizaron en glutaraldeido ácido al 2% por 10 horas y los otros 6 se esterilizaron en calor seco con el horno Dentronix®, a una temperatura de 235 ºC durante veinte minutos (T1). De esta forma se efectuaron tres ciclos mas de esterilización (T2 aT4), posteriores a la activación de 2cm del resorte durante 20 días, simulando una situación clínica. Las graficas de T0 a T4 no mostraron las características clásicas de aleaciones superelásticas. Mediante el ANOVA de dos vías, se determinó que el tipo de esterilización y el número de ciclos de reciclado son variables significativas en el comportamiento mecánico de los resortes. Se generó un modelo, a partir del análisis de regresión múltiple que permitió calcular la variación en la fuerza entregada por los resortes bajo condiciones particulares de tipo de esterilización número de ciclos de reciclado. 


PALABRAS CLAVES

Resortes de Nitinol, Superelasticidad, tipos de reciclaje, grafica tensión/deformación.


ABSTRACT

The objective of this study was to describe the changes in the superelastic properties of the nitinol coil spring exposed to several recycling periods. 12 nitinol coil spring 3MUnitek® of 9mm long were taken in order to determine their behavior into the load/deflection slope (T0). Then they were divided into 2 groups, 6 coil spring were sterilized with acid glutaraldehyde 2% for a period of 10 hours and the other 6 coil spring were sterilized with dry heat into the Dentronix® oven, at 235 ºC during 20min (T1). Following this procedure three more periods of sterilization were done (T2 to T4), after 2cm activation of each coil spring during a period of 20 days, to simulate a clinical situation. The slopes of T0 to T4 did not show the classical characteristics of the superelastics alloys. Through the two way ANOVA analysis it was determined that the method of sterilization and the number of recycling periods are significative characteristics into the coil spring mechanic behavior. A model was generated from the multiple regression analysis that allows calculating the variation in the straight given by the coil spring under particular conditions of sterilization method and number of periods of sterilization.
 

KEY WORDS

Nitinol Coil Spring, superelasticity, methods of sterilization, load/deflection slope

 


Articulo correspondiente al trabajo de grado para optar por el titulo de especialistas en ortodoncia.

Odontóloga, Universidad Nacional. Ortodoncista Pontificia Universidad Javeriana

Odontóloga, Universidad del Zulia. Ortodoncista Pontificia Universidad Javeriana

Odontólogo, Universidad Nacional. Ortodoncista Universidad de Antioquia